63 / 67 page ![]() CS5484 DS981F3 63 7.1.1.2 AC Offset Calibration The AC offset calibration command measures the residual RMS values on the current channel at zero input and stores the squared result in the associated AC offset register. This AC offset will be subtracted from RMS measurements in subsequent conversions, removing the AC offset on the associated current channel. The AC offset register for the channel being calibrated should first be cleared prior to performing the calibration. The high-pass filter should be enabled if AC offset calibration is used. It is recommended that TSETTLE be set to 2000 ms before performing an AC offset calibration. Note that the AC offset register holds the square of RMS value measured during calibration. Therefore, it can hold a maximum RMS noise of . This is the maximum RMS noise that AC offset correction can remove. 7.1.2 Gain Calibration Prior to executing the gain calibration command, gain registers for any path to be calibrated (VxGAIN, IxGAIN) should be set to 1.0, and TSETTLE should be set to 2000 ms. For gain calibration, a reference signal must be applied to the meter. During gain calibration, the voltage RMS result register (VxRMS) is divided into 0.6, and the current RMS result register (IxRMS) is divided into the Scale register. The quotient is put into the associated gain register. The gain calibration algorithm attempts to adjust the gain register (VxGAIN, IxGAIN) such that the voltage RMS result register (VxRMS) equals 0.6, and the current RMS result register (IxRMS) equals the Scale register. Note that for the gain calibration, there are limitations on choosing the reference level and the Scale register value. Using a reference or a scale that is too large or too small can cause register overflow during calibration or later during normal operation. Either condition can set Status register bits I1OR (I2OR) V1OR (V2OR). The maximum value that the gain register can attain is four. Using inappropriate reference levels or scale values may also cause the CS5484 to attempt to set the gain register higher than four, therefore the gain calibration result will be invalid. The Scale register is 0.6 by default. The maximum voltage (UMAX Volts) and current (IMAX Amps) of the meter should be used as the reference signal level if the Scale register is 0.6. After gain calibration, 0.6 of the VxRMS (IxRMS) register represents UMAX Volts (IMAX Amps) for the line voltage (load current); 0.36 of the PxAVG, QxAVG, or Sx register represents UMAX×IMAX Watts, Vars, or VAs for the active, reactive, or apparent power. If the calibration is performed with UMAX Volts and ICAL Amps and ICAL<IMAX, the Scale register needs to be scaled down to 0.6×ICAL/IMAX before performing gain calibration. After gain calibration, 0.6 of the VxRMS register represents UMAX Volts, 0.6xICAL/IMAX of the IxRMS register represents ICAL Amps, and 0.36 x ICAL /IMAX of the PxAVG, QxAVG, or Sx register represents UMAXxICAL Watts, Vars, or VAs. 7.1.3 Calibration Order 1. If the HPF option is enabled, then any DC component that may be present in the selected signal channel will be removed, and a DC offset calibration is not required. However, if the HPF option is disabled, the DC offset calibration should be performed. When using high-pass filters, it is recommended that the DC offset register for the corresponding channel be set to zero. Before performing DC offset calibra- tion, the DC offset register should be set to 0, and the corresponding gain register should be set to 1. 2. If there is an AC offset in the IxRMS calculation, the AC offset calibration should be performed on the current channel. Before performing AC offset calibration, the AC offset register should be set to 0. It is recommended that TSETTLE be set to 2000ms before performing an AC offset calibration. 3. Perform the gain calibration. 4. If an AC offset calibration was performed (step 2), then the AC offset may need to be adjusted to compensate for the change in gain (step 3). This can be accomplished by restoring zero to the AC offset register and then performing an AC offset calibration. The adjustment could also be done by multiplying the AC offset register value that was calculated in step 2 by the gain calculated in step 3 and updating the AC offset register with the product. 7.2 Phase Compensation A phase compensation mechanism is provided to adjust for meter-to-meter variation in signal path delays. Phase offset between a voltage channel and its corresponding current channel can be calculated by using the power factor (PF1, PF2) register after a conversion. 1) Apply a reference voltage and current with a lagging power factor to the meter. The reference current waveform should lag the voltage with a 60° phase shift. 2) Start continuous conversion. 0xFFFFFF |
|
Link URL |
| Cho đến nay ALLDATASHEET có giúp ích cho doanh nghiệp của bạn hay không? [ DONATE ] |
Alldatasheet là | Quảng cáo | Liên lạc với chúng tôi | Chính sách bảo mật | Liên kết đến bảng dữ liệu | Trao đổi link | Tìm kiếm theo nhà sản xuất All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |