28 / 52 page ![]() 4.5 Analog Input Block The ANALOG INPUT (AI) block takes the Transducer Block input data, selected by channel number, and makes it available to other function blocks at its output: 1. Level 2. Volume 3. Interface 4. Interface Volume 4.5.1 AI Block Parameters The first eight parameters in an AI block are the standard block parameters discussed in section 4.1.1. Additional analog input function block parameters are as follows: BATCH: A parameter intended to be used in Batch applica- tion in line with IEC 61512 Part 1. OUT: Contains the current measurement value in the con- figuration engineering unit. PV_SCALE: High and low scale values used to convert Process Variable Configured by channel into percent. OUT_SCALE: The high and low scale values, the engineer- ing code, and number of digits to the right of the decimal point to be used in displaying the OUT parameters. LIN_TYPE: Type of linearization. No linerization is sup- ported in the AI function blocks. CHANNEL: Selects the measurement value from an active transducer block as the input to the function block. PV_FTIME: Filter time of the Process Variable. FSAFE_TYPE: Defines the reaction of a device, if a fault is detected and the quality of the process variable input from the transducer block is BAD. 0 = FSAFE_VALUE is used as OUT 1 = Use last stored valid OUT value 2 = OUT has incorrect calculated value; status remains “bad”. FSAFE_VALUE: Default value for the OUT parameter, if a fault is detected, and FSAFE_TYPE is 0. HI_LIM: Value of the upper limit of warnings. LO_LIM: Value of the lower limit of warnings. ALARM_HYS: Hysteresis to adjust sensitivity of alarm triggering. HI_HI_LIM: Value of the upper limit of alarms. HI_LIM: Value of the upper limit of warnings. LO_LIM: Value of the lower limit of warnings. 28 57-645 Eclipse Guided Wave Radar Transmitter - PROFIBUS PA™ Scaling |
|
Link URL |
| Cho đến nay ALLDATASHEET có giúp ích cho doanh nghiệp của bạn hay không? [ DONATE ] |
Alldatasheet là | Quảng cáo | Liên lạc với chúng tôi | Chính sách bảo mật | Liên kết đến bảng dữ liệu | Trao đổi link | Tìm kiếm theo nhà sản xuất All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |