Back Button ADIS16475-1/PCBZ Datasheet HTML 8Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 9Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 10Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 11Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 12Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 13Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 14Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 15Page - Analog Devices ADIS16475-1/PCBZ Datasheet HTML 16Page - Analog Devices Next Button
Zoom Inzoom in Zoom Outzoom out
 12 / 37 page
background image
ADIS16475
Data Sheet
Rev. D | Page 12 of 37
THEORY OF OPERATION
INTRODUCTION
When using the factory default configuration for all user
configurable control registers, the ADIS16475 initializes itself
and automatically starts a continuous process of sampling,
processing, and loading calibrated sensor data into its output
registers at a rate of 2000 SPS.
INERTIAL SENSOR SIGNAL CHAIN
Figure 18 provides the basic signal chain for the inertial sensors
in the ADIS16475. This signal chain produces an update rate
of 2000 SPS in the output data registers when it operates in
internal clock mode (default, see Register MSC_CTRL, Bits[4:2]
in Table 105).
OUTPUT
DATA
REGISTERS
AVERAGING
DECIMATING
FILTER
MEMS
SENSORS
CALIBRATION
BARTLETT
WINDOW
FIR
FILTER
Figure 18. Signal Processing Diagram, Inertial Sensors
Gyroscope Data Sampling
The three gyroscopes produce angular rate measurements around
three orthogonal axes (x, y, and z). Figure 19 shows the data
sampling plan for each gyroscope when the ADIS16475 operates
in internal clock mode (default, see Register MSC_CTRL,
Bits[4:2] in Table 105). Each gyroscope has an analog-to-digital
converter (ADC) and sample clock (fSG) that drives data sampling
at a rate of 4100 Hz (±5%). The internal processor reads and
processes this data from each gyroscope at a rate of 2000 Hz (fSM).
MEMS
GYROSCOPE
fSM = 2000Hz
INTERNAL
DATA
REGISTER
fSG = 4100Hz
ADC
TO
BARTLETT
WINDOW
FIR FILTER
Figure 19. Gyroscope Data Sampling
Accelerometer Data Sampling
The three accelerometers produce linear acceleration measurements
along the same orthogonal axes (x, y, and z) as the gyroscopes.
Figure 20 shows the data sampling plan for each accelerometer
when the ADIS16475 operates in internal clock mode (default,
see Register MSC_CTRL, Bits[4:2] in Table 105).
MEMS
ACCELEROMETER
2 ×
fSM = 4000Hz
ADC
1
2
2
n = 1
a(n)
÷2
TO
BARTLETT
WINDOW
FIR FILTER
Figure 20. Accelerometer Data Sampling
External Clock Options
The ADIS16475 provides three different modes of operation
that support the device using an external clock to control the
internal processing rate (fSM in Figure 19 and Figure 20) through
the SYNC pin. The MSC_CTRL register (see Table 105) provides
the configuration options for these external clock modes in
Bits[4:2].
Inertial Sensor Calibration
The inertial sensor calibration function for the gyroscopes and the
accelerometers has two components: factory calibration and
user calibration (see Figure 21).
TO
AVERAGING
DECIMATING
FILTER
FACTORY
CALIBRATION
USER
CALIBRATION
FROM
BARTLETT
WINDOW
FIR FILTER
Figure 21. Inertial Sensor Calibration Processing
The factory calibration of the gyroscope applies the following
correction formulas to the data of each gyroscope:
×
+
+
×
=
ZC
YC
XC
33
32
31
23
22
21
13
12
11
Z
Y
X
Z
Y
X
33
32
31
23
22
21
13
12
11
ZC
YC
XC
a
a
a
l
l
l
l
l
l
l
l
l
b
b
b
ω
ω
ω
m
m
m
m
m
m
m
m
m
ω
ω
ω
where:
ωXC, ωYC, and ωZC are the gyroscope outputs (post calibration).
m11, m12, m13, m21, m22, m23, m31, m32, and m33 provide scale and
alignment correction.
ωX, ωY, and ωZ are the gyroscope outputs (precalibration).
bX, bY, and bZ provide bias correction.
l11, l12, l13, l21, l22, l23, l31, l32, and l33 provide linear g correction
aXC, aYC, and aZC are the accelerometer outputs (post calibration).
All of the correction factors in this relationship come from
direct observation of the response of each gyroscope at multiple
temperatures over the calibration temperature range (−40°C ≤
TC ≤ +85°C). These correction factors are stored in the flash
memory bank, but they are not available for observation or
configuration. Register MSC_CTRL, Bit 7 (see Table 105)
provides the only user configuration option for the factory
calibration of the gyroscopes: an on/off control for the linear g
compensation. See Figure 44 for more details on the user
calibration options available for the gyroscopes.



Html Pages

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37


bảng dữ liệu tải về

Go To PDF Page


Link URL



Cho đến nay ALLDATASHEET có giúp ích cho doanh nghiệp của bạn hay không?  [ DONATE ] 

Alldatasheet là   |   Quảng cáo   |   Liên lạc với chúng tôi   |   Chính sách bảo mật   |   Liên kết đến bảng dữ liệu    |   Trao đổi link   |   Tìm kiếm theo nhà sản xuất
All Rights Reserved©Alldatasheet.com


Mirror Sites
English : Alldatasheet.com  |   English : Alldatasheet.net  |   Chinese : Alldatasheetcn.com  |   German : Alldatasheetde.com  |   Japanese : Alldatasheet.jp
Russian : Alldatasheetru.com  |   Korean : Alldatasheet.co.kr  |   Spanish : Alldatasheet.es  |   French : Alldatasheet.fr  |   Italian : Alldatasheetit.com
Portuguese : Alldatasheetpt.com  |   Polish : Alldatasheet.pl  |   Vietnamese : Alldatasheet.vn
Indian : Alldatasheet.in  |   Mexican : Alldatasheet.com.mx  |   British : Alldatasheet.co.uk  |   New Zealand : Alldatasheet.co.nz
Family Site : ic2ic.com  |   icmetro.com