M2RS-3E-R2/CE
AI

The **M2RS-3E-R2/CE** is a high-precision, plug-in type **RTD (Resistance Temperature Detector) Transmitter** manufactured by **M-System**. It is designed to accept a signal from an RTD sensor and convert it into a standard linear DC output signal.
Below is a breakdown of its electronic specifications and characteristics.
---
## 1. Core Specifications
The model code `M2RS-3E-R2/CE` identifies specific electronic parameters:
| Component | Specification Detail | Description |
| :--- | :--- | :--- |
| **Model Series** | M2RS | Super-mini Terminal Block Type Signal Conditioner |
| **Input (3)** | JPt 100 (JIS '89) | Japanese standard platinum RTD (0 to 500°C range typically) |
| **Output (E)** | 0 – 5V DC | Voltage output signal |
| **Power Input (R2)** | 11 – 27V DC | Standard DC operating voltage |
| **Standards (/CE)** | CE Marking | Compliant with European safety and EMC standards |
---
## 2. Electronic Components & Internal Functionality
The device operates through a series of internal electronic stages to ensure signal integrity:
### A. Input Stage (RTD Interface)
* **Sensor Connection:** 3-wire system. This is crucial for compensating for lead-wire resistance, ensuring the electronic measurement reflects only the temperature at the sensor tip.
* **Excitation Current:** The unit provides a small, constant sensing current (typically around 1mA) to the RTD to prevent self-heating while measuring resistance.
### B. Signal Conditioning (Amplification & Linearization)
* **Linearization Circuitry:** RTDs have a slightly non-linear temperature-to-resistance relationship. The M2RS includes electronic linearization to ensure the 0–5V output is perfectly proportional to the temperature.
* **Zero/Span Adjustments:** Located on the front panel, these allow for fine-tuning of the electronic output using a multi-turn potentiometer.
### C. Isolation Technology
* **Three-Way Isolation:** The device provides electronic isolation between the **Input**, **Output**, and **Power** terminals.
* **Isolation Resistance:** Typically 100 MΩ or more with 500V DC.
* **Dielectric Strength:** Capable of withstanding 2000V AC for 1 minute (Input to Output to Power to Ground).
---
## 3. Performance Parameters
| Parameter | Value |
| :--- | :--- |
| **Accuracy** | ±0.2% of span |
| **Temperature Coefficient** | ±0.015% / °C (±0.008% / °F) |
| **Response Time** | ≤ 0.5 seconds (0 – 90%) |
| **Burnout Protection** | Standard (Output goes to Upscale in case of wire break) |
| **Power Consumption** | Approx. 0.8W |
---
## 4. Physical Connection Diagram
The electronic terminals are arranged for DIN-rail mounting or surface mounting:
```text
Terminal | Function
---------|---------------------------
1 & 2 | Output (+) / (-)
4 & 5 | Power Supply (+) / (-)
A, B, B'| RTD Input (3-wire connection)
```
---
- ⤷
How do I calibrate the zero and span settings for the M2RS-3E-R2?
- ⤷ What are the specific wiring requirements for the 3-wire RTD connection to avoid interference?
- ⤷ Can this model be used with a Pt100 (IEC) sensor instead of the JPt100?